Table Top Steam Autoclaves/Sterilizers
Vertical Autoclaves / Sterilizers
Large Pulsant Vacuum Autoclaves /Sterilizers
Hot Air Autoclaves / Sterilizers
 
GMQ.C Series Fully Automatic Hot Air Autoclaves/Sterilizers
DHG Series Heating and Drying Oven
GRX Series Hot Air Sterilizer
GZX Series Hot Air Sterilizer
Washer Disinfectors
Table Top Ultrasonic Cleaners
Industrial Ultrasonic Cleaners
Accessories
New Products


Products>>Table-Top Steam Autoclaves/Sterilizers

   
CLASS N-TMQ.C 5090 185L
TMQ.C Series Fully Automatic Table Top Autoclaves

Application:
It is suitable to be used in sterilizing dressing, Class B Cavity devices, solid instruments, medical catheters, culture medium, and waste in the operating room, clinic department, CSSD, biological laboratories and other healthcare sectors.

Main Features:
●Microcomputer control technology, touch screen;
●Affusion, Heating, sterilization, exhaust, drying whole processes are automatically controlled by program;
●LCD dynamically display process,state, temperature, time;
●Dynamic pulse with a positive-pressure exhaust functions, completely exhausting cold air in the cavity, to ensure the steam saturation;
●With instruments, instruments bags, dressings, rubber, liquid and other sterilization procedures for customized optional;
●With evaporator, external water;
●Self-check and LCD display mulfunction;
●Beep reminding when ending sterilization;
●Mini printer is optional which record the whole sterilization process.

Safety Protection:
●Over-temperature protection: Over setting temperature, system automatically cut off power;
●Door safety interlock devices: the door is only closed in place, power can be connected to heat steam; If there is pressure inside the chamber, the door will not be opened;
●Automatic releasing pressure safety valve: Over setting pressure, releasing valve opens to release the pressure;
●Anti-dry alarm: below the water level, automatically cut off the heating power, and display alarm;
● Electronic circuit safety devices: AC main circuit with a short-circuit protection, over-voltage DC control circuits with overload protection.

Main Technical Parameters:
Designed Pressure: 0.28MPa
Designed Temperature: 150℃
Max. Working Pressure: 0.23MPa
Temperature Selective Range: 105℃~136℃
Temperature Precision: ±0.1℃
Temperature uniformity: ≦±0.5℃
Chamber Material: SUS304 stainless steel

   



Contact Information
Yongfeng Enterprise Co., Ltd.
Add: No.1, Lane 110, Shengfeng Road, Ningbo, 315010, China
Tel:  86-574-87255354
Fax: 86-574-87125894
Web Site: http://www.china-sterilizers.com, http://autoclave.chinese-suppliers.com
E-mail: info@china-sterilizers.com

Copyright 2015, All rights reserved. Designed & Supported By Haibo Network Inc.